Publications & Posters

Highly specific and ultrasensitive plasma test detects Abeta(1-42) and Abeta(1-40) in Alzheimer’s disease

Scientific Reports | May 6, 2021

Thijssen EH, Verberk IMW, Vanbrabant J, Koelewijn A, Heijst H, Scheltens P, van der Flier W, Vanderstichele H, Stoops E and Teunissen CE

Sci Rep. 2021;11:9736

DOi: https://doi.org/10.1038/s41598-021-89004-x

This study was performed using a Simoa Homebrew assay for Aβ40 and Aβ42.

Abstract

Plasma biomarkers that reflect specific amyloid beta (Abeta) proteoforms provide an insight in the treatment effects of Alzheimer’s disease (AD) therapies. Our aim was to develop and validate ready-to-use Simoa ‘Amyblood’ assays that measure full length Abeta1-42 and Abeta1-40 and compare their performance with two commercial assays. Linearity, intra- and inter-assay %CV were compared between Amyblood, Quanterix Simoa triplex, and Euroimmun ELISA. Sensitivity and selectivity were assessed for Amyblood and the Quanterix triplex. Clinical performance was assessed in CSF biomarker confirmed AD (n = 43, 68 ± 6 years) and controls (n = 42, 62 ± 5 years). Prototype and Amyblood showed similar calibrator curves and differentiation (20 AD vs 20 controls, p < 0.001). Amyblood, Quanterix triplex, and ELISA showed similar linearity (96%-122%) and intra-assay %CVs (≤ 3.1%). A minor non-specific signal was measured with Amyblood of + 2.4 pg/mL Abeta1-42 when incubated with 60 pg/mL Abeta1-40. A substantial non-specific signal of + 24.7 pg/mL Abetax-42 was obtained when 40 pg/mL Abeta3-42 was measured with the Quanterix triplex. Selectivity for Abeta1-42 at physiological Abeta1-42 and Abeta1-40 concentrations was 125% for Amyblood and 163% for Quanterix. Amyblood and Quanterix ratios (p < 0.001) and ELISA Abeta1-42 concentration (p = 0.025) could differentiate AD from controls. We successfully developed and upscaled a prototype to the Amyblood assays with similar technical and clinical performance as the Quanterix triplex and ELISA, but better specificity and selectivity than the Quanterix triplex assay. These results suggest leverage of this specific assay for monitoring treatment response in trials.

This study was performed using the Quanterix HD-1 Analyzer.