Head-to-head validation of six immunoassays for SARS-CoV-2 in hospitalized patients

Journal of Clinical Virology | April 18, 2021

Zonneveld R, Jurriaans S, van Gool T, Hofstra JJ, Hekker TAM, Defoer P, Broekhuizen-van Haaften PE, Wentink-Bonnema EM, Boonkamp L, Teunissen CE, Heijboer AC, Martens F, de Bree G, van Vugt M and van Houdt R

J Clin Virol. 2021 Apr 18;139:104821

DOI: https://doi.org/10.1016/j.jcv.2021.104821

Abstract

Background

Detecting SARS-CoV-2 antibodies may help to diagnose COVID-19. Head-to-head validation of different types of immunoassays in well-characterized cohorts of hospitalized patients remains needed.

Methods

We validated three chemiluminescence immunoassays (CLIAs) (Liaison, Elecsys, and Abbott) and one single molecule array assay (SIMOA) (Quanterix) for automated analyzers, one rapid immunoassay RIA (AllTest), and one ELISA (Wantai) in parallel in first samples from 126 PCR confirmed COVID-19 hospitalized patients and 158 pre-COVID-19 patients. Specificity of the AllTest was also tested in 106 patients with confirmed parasitic and dengue virus infections. Specificity of the Wantai assay was not tested due to limitations in sample volumes.

Results

Overall sensitivity in first samples was 70.6 % for the Liaison, 71.4 % for the Elecsys, 75.4 % for the Abbott, 70.6 % for the Quanterix, 77.8 % for the AllTest, and 88.9 % for the Wantai assay, respectively. Sensitivity was between 77.4 % (Liaison) and 94.0 % (Wantai) after 10 dpso. No false positive results were observed for the Elecsys and Abbott assays. Specificity was 91.1 % for the Quanterix, 96.2 % for the Liaison, and 98.1 % for the AllTest assay, respectively.

Conclusion

We conclude that low sensitivity of all immunoassays limits their use early after onset of illness in diagnosing COVID-19 in hospitalized patients. After 10 dpso, the Wantai ELISA has a relatively high sensitivity, followed by the point-of-care AllTest RIA that compares favorably with automated analyzer immunoassays.